Features:
1. The large sample chamber can accommodate 5-100mm cuvettes.
2. It can be directly connected to a dedicated printer (optional) to print data.
3. The key components are all imported with original packaging to ensure the accuracy and precision of the measurement.
4. Digital display of measured value.
5. The instrument adopts advanced microcomputer processing technology and is easy to operate.
6. The automatic light door guarantees the service life of the photoelectric sensor and makes the instrument test easier.
Model | UV752 | UV752N |
Optical System | Czerny-Turner (1200L/mm grating) |
Wavelength Range | 195-1020 nm | 195-1020 nm |
Spectral Bandwidth | 5 nm | 4nm |
Wavelength Accuracy | ±2nm | ±2nm |
Wavelength Repeatability | 1nm |
Photometric Accuracy | ±0.5%T | ±0.5%T |
Photometric Repeatability | 0.2%T | 0.2%T |
Photometric Range | 0-200%T, -0.301A~2.000A |
Stray light | 0.2%T | 0.1%T |
Stability | 0.004 A/h@500nm | 0.002A/h@500nm |
Baseline Flatness | 0.004 A/h@500nm | 0.002A/h@500nm |
Noise | 100%≤0.3T |
Wavelength Setting | Manual |
Work Mode | T, A, C |
Light Source | Deuterium &Tungsten lamp | Deuterium &Tungsten lamp |
Display | 4 inch LCD |
Detector | Silicon Photodiode |
Output | RS232 port/Printer port(optional) |
Power | AC 220V/50Hz or AC 110V/60Hz |
Shipping Size | 520*450*320mm/12kg | 520*450*320mm/12kg |